This site uses cookies, tags, and tracking settings to store information that help give you the very best browsing experience. Dismiss this warning

Safety and efficacy of the Pipeline Embolization Device in 100 small intracranial aneurysms

Nohra Chalouhi Department of Neurological Surgery, Thomas Jefferson University and Jefferson Hospital for Neuroscience, Philadelphia, Pennsylvania;

Search for other papers by Nohra Chalouhi in
Current site
Google Scholar
PubMed
Close
 MD
,
Mario Zanaty Department of Neurological Surgery, Thomas Jefferson University and Jefferson Hospital for Neuroscience, Philadelphia, Pennsylvania;

Search for other papers by Mario Zanaty in
Current site
Google Scholar
PubMed
Close
 MD
,
Alex Whiting Department of Neurological Surgery, Thomas Jefferson University and Jefferson Hospital for Neuroscience, Philadelphia, Pennsylvania;

Search for other papers by Alex Whiting in
Current site
Google Scholar
PubMed
Close
 MD
,
Steven Yang Department of Neurological Surgery, Thomas Jefferson University and Jefferson Hospital for Neuroscience, Philadelphia, Pennsylvania;

Search for other papers by Steven Yang in
Current site
Google Scholar
PubMed
Close
 MS
,
Stavropoula Tjoumakaris Department of Neurological Surgery, Thomas Jefferson University and Jefferson Hospital for Neuroscience, Philadelphia, Pennsylvania;

Search for other papers by Stavropoula Tjoumakaris in
Current site
Google Scholar
PubMed
Close
 MD
,
David Hasan Department of Neurosurgery, University of Iowa, Iowa City, Iowa; and

Search for other papers by David Hasan in
Current site
Google Scholar
PubMed
Close
 MD
,
Robert M. Starke Department of Neurological Surgery, University of Virginia, Charlottesville, Virginia

Search for other papers by Robert M. Starke in
Current site
Google Scholar
PubMed
Close
 MD
,
Shannon Hann Department of Neurological Surgery, Thomas Jefferson University and Jefferson Hospital for Neuroscience, Philadelphia, Pennsylvania;

Search for other papers by Shannon Hann in
Current site
Google Scholar
PubMed
Close
 MD
,
Christine Hammer Department of Neurological Surgery, Thomas Jefferson University and Jefferson Hospital for Neuroscience, Philadelphia, Pennsylvania;

Search for other papers by Christine Hammer in
Current site
Google Scholar
PubMed
Close
 MD
,
David Kung Department of Neurological Surgery, Thomas Jefferson University and Jefferson Hospital for Neuroscience, Philadelphia, Pennsylvania;

Search for other papers by David Kung in
Current site
Google Scholar
PubMed
Close
 MD
,
Robert Rosenwasser Department of Neurological Surgery, Thomas Jefferson University and Jefferson Hospital for Neuroscience, Philadelphia, Pennsylvania;

Search for other papers by Robert Rosenwasser in
Current site
Google Scholar
PubMed
Close
 MD
, and
Pascal Jabbour Department of Neurological Surgery, Thomas Jefferson University and Jefferson Hospital for Neuroscience, Philadelphia, Pennsylvania;

Search for other papers by Pascal Jabbour in
Current site
Google Scholar
PubMed
Close
 MD
Restricted access

Purchase Now

USD  $45.00

JNS + Pediatrics - 1 year subscription bundle (Individuals Only)

USD  $536.00

JNS + Pediatrics + Spine - 1 year subscription bundle (Individuals Only)

USD  $636.00
USD  $45.00
USD  $536.00
USD  $636.00
Print or Print + Online Sign in

OBJECT

Flow diverters are increasingly used for treatment of intracranial aneurysms. In most series, the Pipeline Embolization Device (PED) was used for the treatment of large, giant, complex, and fusiform aneurysms. Little is known about the use of the PED in small aneurysms. The purpose of this study was to assess the safety and efficacy of the PED in small aneurysms (≤ 7 mm).

METHODS

A total of 100 consecutive patients were treated with the PED at the authors' institution between May 2011 and September 2013. Data on procedural safety and efficacy were retrospectively collected.

RESULTS

The mean aneurysm size was 5.2 ± 1.5 mm. Seven patients (7%) had sustained a subarachnoid hemorrhage. All except 5 aneurysms (95%) arose from the anterior circulation. The number of PEDs used was 1.2 per aneurysm. Symptomatic procedure-related complications occurred in 3 patients (3%): 1 distal parenchymal hemorrhage that was managed conservatively and 2 ischemic events. At the latest follow-up (mean 6.3 months), 54 (72%) aneurysms were completely occluded (100%), 10 (13%) were nearly completely occluded (≥ 90%), and 11 (15%) were incompletely occluded (< 90%). Six aneurysms (8%) required further treatment. Increasing aneurysm size (OR 3.8, 95% CI 0.99–14; p = 0.05) predicted retreatment. All patients achieved a favorable outcome (modified Rankin Scale Score 0–2) at follow-up.

CONCLUSIONS

In this study, treatment of small aneurysms with the PED was associated with low complication rates and high aneurysm occlusion rates. These findings suggest that the PED is a safe and effective alternative to conventional endovascular techniques for small aneurysms. Randomized trials with long-term follow-up are necessary to determine the optimal treatment that leads to the highest rate of obliteration and the best clinical outcomes.

ABBREVIATIONS

ICA = internal carotid artery; MCA = middle cerebral artery; mRS = modified Rankin Scale; PED = Pipeline Embolization Device; SAH = subarachnoid hemorrhage.
  • Collapse
  • Expand
  • 1

    Becske T, , Kallmes DF, , Saatci I, , McDougall CG, , Szikora I, & Lanzino G, et al.: Pipeline for uncoilable or failed aneurysms: results from a multicenter clinical trial. Radiology 267:858868, 2013

    • Crossref
    • PubMed
    • Search Google Scholar
    • Export Citation
  • 2

    Chalouhi N, , Dumont AS, , Randazzo C, , Tjoumakaris S, , Gonzalez LF, & Rosenwasser R, et al.: Management of incidentally discovered intracranial vascular abnormalities. Neurosurg Focus 31:6 E1, 2011

    • Crossref
    • PubMed
    • Search Google Scholar
    • Export Citation
  • 3

    Chalouhi N, , Jabbour P, , Singhal S, , Drueding R, , Starke RM, & Dalyai RT, et al.: Stent-assisted coiling of intracranial aneurysms: predictors of complications, recanalization, and outcome in 508 cases. Stroke 44:13481353, 2013

    • Crossref
    • PubMed
    • Search Google Scholar
    • Export Citation
  • 4

    Chalouhi N, , Satti SR, , Tjoumakaris S, , Dumont AS, , Gonzalez LF, & Rosenwasser R, et al.: Delayed migration of a Pipeline Embolization Device. Neurosurgery 72:2 Suppl Operative ons229ons234, 2013

    • PubMed
    • Search Google Scholar
    • Export Citation
  • 5

    Chalouhi N, , Tjoumakaris S, , Starke RM, , Gonzalez LF, , Randazzo C, & Hasan D, et al.: Comparison of flow diversion and coiling in large unruptured intracranial saccular aneurysms. Stroke 44:21502154, 2013

    • Crossref
    • PubMed
    • Search Google Scholar
    • Export Citation
  • 6

    Chalouhi N, , Tjoumakaris SI, , Gonzalez LF, , Hasan D, , Pema PJ, & Gould G, et al.: Spontaneous delayed migration/shortening of the pipeline embolization device: report of 5 cases. AJNR Am J Neuroradiol 34:23262330, 2013

    • Crossref
    • PubMed
    • Search Google Scholar
    • Export Citation
  • 7

    D'Urso PI, , Lanzino G, , Cloft HJ, & Kallmes DF: Flow diversion for intracranial aneurysms: a review. Stroke 42:23632368, 2011

  • 8

    De Vries J, , Boogaarts J, , Van Norden A, & Wakhloo AK: New generation of Flow Diverter (surpass) for unruptured intracranial aneurysms: a prospective single-center study in 37 patients. Stroke 44:15671577, 2013

    • Crossref
    • PubMed
    • Search Google Scholar
    • Export Citation
  • 9

    Deshmukh V, , Hu YC, , McDougall CG, , Barnwell SL, , Albuquerque F, & Fiorella D: Histopathological assessment of delayed ipsilateral parenchymal hemorrhages after the treatment of paraclinoid aneurysms with the Pipeline Embolization Device. Neurosurgery 71:E551E552, 2012

    • Crossref
    • PubMed
    • Search Google Scholar
    • Export Citation
  • 10

    Morita A, , Kirino T, , Hashi K, , Aoki N, , Fukuhara S, & Hashimoto N, et al.: The natural course of unruptured cerebral aneurysms in a Japanese cohort. N Engl J Med 366:24742482, 2012

    • Crossref
    • PubMed
    • Search Google Scholar
    • Export Citation
  • 11

    Jabbour P, , Chalouhi N, , Tjoumakaris S, , Gonzalez LF, , Dumont AS, & Randazzo C, et al.: The Pipeline Embolization Device: learning curve and predictors of complications and aneurysm obliteration. Neurosurgery 73:113120, 2013

    • Crossref
    • PubMed
    • Search Google Scholar
    • Export Citation
  • 12

    Jabbour PM, , Chalouhi N, & Rosenwasser RH: The Pipeline Embolization Device: what have we learned?. World Neurosurg 80:798799, 2013

  • 13

    Kallmes DF, , Ding YH, , Dai D, , Kadirvel R, , Lewis DA, & Cloft HJ: A new endoluminal, flow-disrupting device for treatment of saccular aneurysms. Stroke 38:23462352, 2007

    • Crossref
    • PubMed
    • Search Google Scholar
    • Export Citation
  • 14

    Kan P, , Siddiqui AH, , Veznedaroglu E, , Liebman KM, , Binning MJ, & Dumont TM, et al.: Early postmarket results after treatment of intracranial aneurysms with the pipeline embolization device: a U.S. multicenter experience. Neurosurgery 71:10801088, 2012

    • Crossref
    • PubMed
    • Search Google Scholar
    • Export Citation
  • 15

    Lin LM, , Colby GP, , Kim JE, , Huang J, , Tamargo RJ, & Coon AL: Immediate and follow-up results for 44 consecutive cases of small (<10 mm) internal carotid artery aneurysms treated with the pipeline embolization device. Surg Neurol Int 4:114, 2013

    • Crossref
    • PubMed
    • Search Google Scholar
    • Export Citation
  • 16

    Piano M, , Valvassori L, , Quilici L, , Pero G, & Boccardi E: Mid-term and long-term follow-up of cerebral aneurysms treated with flow diverter devices: a single-center experience. J Neurosurg 118:408416, 2013

    • Crossref
    • PubMed
    • Search Google Scholar
    • Export Citation
  • 17

    Pierot L, & Wakhloo AK: Endovascular treatment of intracranial aneurysms: current status. Stroke 44:20462054, 2013

  • 18

    Pistocchi S, , Blanc R, , Bartolini B, & Piotin M: Flow diverters at and beyond the level of the circle of Willis for the treatment of intracranial aneurysms. Stroke 43:10321038, 2012

    • Crossref
    • PubMed
    • Search Google Scholar
    • Export Citation
  • 19

    Saatci I, , Yavuz K, , Ozer C, , Geyik S, & Cekirge HS: Treatment of intracranial aneurysms using the Pipeline Flow-Diverter Embolization Device: a single-center experience with longterm follow-up results. AJNR Am J Neuroradiol 33:14361446, 2012

    • Crossref
    • PubMed
    • Search Google Scholar
    • Export Citation
  • 20

    Yu SC, , Kwok CK, , Cheng PW, , Chan KY, , Lau SS, & Lui WM, et al.: Intracranial aneurysms: midterm outcome of Pipeline Embolization Device—a prospective study in 143 patients with 178 aneurysms. Radiology 265:893901, 2012

    • Crossref
    • PubMed
    • Search Google Scholar
    • Export Citation

Metrics

All Time Past Year Past 30 Days
Abstract Views 372 92 49
Full Text Views 1438 244 2
PDF Downloads 1115 170 3
EPUB Downloads 0 0 0