In thoracolumbar deformity surgery, anterior-only approaches are used for reconstruction of anterior column failures. It is generally advised that vertebral body replacements (VBRs) should be preloaded by compression. However, little is known regarding the impact of different techniques for generation of preloads and which surgical principle is best for restoration of lordosis. Therefore, the authors analyzed the effect of different surgical techniques to restore spinal alignment and lordosis as well as the ability to generate axial preloads on VBRs in anterior column reconstructions.
The authors performed a laboratory study using 7 fresh-frozen specimens (from T-3 to S-1) to assess the ability for lordosis reconstruction of 5 techniques and their potential for increasing preloads on a modified distractable VBR in a 1-level thoracolumbar corpectomy. The testing protocol was as follows: 1) Radiographs of specimens were obtained. 2) A 1-level corpectomy was performed. 3) In alternating order, lordosis was applied using 1 of the 5 techniques. Then, preloads during insertion and after relaxation using the modified distractable VBR were assessed using a miniature load-cell incorporated in the modified distractable VBR. The modified distractable VBR was inserted into the corpectomy defect after lordosis was applied using 1) a lamina spreader; 2) the modified distractable VBR only; 3) the ArcoFix System (an angular stable plate system enabling in situ reduction); 4) a lordosizer (a customized instrument enabling reduction while replicating the intervertebral center of rotation [COR] according to the COR method); and 5) a lordosizer and top-loading screws ([LZ+TLS], distraction with the lordosizer applied on a 5.5-mm rod linked to 2 top-loading pedicle screws inserted laterally into the vertebra). Changes in the regional kyphosis angle were assessed radiographically using the Cobb method.
The bone mineral density of specimens was 0.72 ± 22.6 g/cm2. The maximum regional kyphosis angle reconstructed among the 5 techniques averaged 9.7°−16.1°, and maximum axial preloads averaged 123.7–179.7 N. Concerning correction, in decreasing order the LZ+TLS, lordosizer, and ArcoFix System outperformed the lamina spreader and modified distractable VBR. The order of median values for insertion peak load, from highest to lowest, were lordosizer, LZ+TLS, and ArcoFix, which outperformed the lamina spreader and modified distractable VBR. In decreasing order, the axial preload was highest with the lordosizer and LZ+TLS, which both outperformed the lamina spreader and the modified distractable VBR. The technique enabling the greatest lordosis achieved the highest preloads. With the ArcoFix System and LZ+TLS, compression loads could be applied and were 247.8 and 190.6 N, respectively, which is significantly higher than the insertion peak load and axial preload (p < 0.05).
Including the ability for replication of the COR in instruments designed for anterior column reconstructions, the ability for lordosis restoration of the anterior column and axial preloads can increase, which in turn might foster fusion.
Please include this information when citing this paper: published online May 18, 2012; DOI: 10.3171/2012.4.SPINE11198.
Acaroglu ER, , Schwab FJ, & Farcy JP: Simultaneous anterior and posterior approaches for correction of late deformity due to thoracolumbar fractures. Eur Spine J 5:56–62, 1996
Akamaru T, , Kawahara N, , Sakamoto J, , Yoshida A, , Murakami H, & Hato T, et al.: The transmission of stress to grafted bone inside a titanium mesh cage used in anterior column reconstruction after total spondylectomy: a finite-element analysis. Spine (Phila Pa 1976) 30:2783–2787, 2005
An HS, , Xu R, , Lim TH, , McGrady L, & Wilson C: Prediction of bone graft strength using dual-energy radiographic absorptiometry. Spine (Phila Pa 1976) 19:2358–2363, 1994
Arts MP, & Peul WC: Vertebral body replacement systems with expandable cages in the treatment of various spinal pathologies: a prospectively followed case series of 60 patients. Neurosurgery 63:537–545, 2008
Bence T, , Schreiber U, , Grupp T, , Steinhauser E, & Mittelmeier W: Two column lesions in the thoracolumbar junction: anterior, posterior or combined approach? A comparative biomechanical in vitro investigation. Eur Spine J 16:813–820, 2007
Benli IT, , Kaya A, , Uruç V, & Akalin S: Minimum 5-year follow-up surgical results of post-traumatic thoracic and lumbar kyphosis treated with anterior instrumentation: comparison of anterior plate and dual rod systems. Spine (Phila Pa 1976) 32:986–994, 2007
Berven SH, , Deviren V, , Smith JA, , Emami A, , Hu SS, & Bradford DS: Management of fixed sagittal plane deformity: results of the transpedicular wedge resection osteotomy. Spine (Phila Pa 1976) 26:2036–2043, 2001
Bolger C, , Bourlion M, , Leroy X, , Petit D, , Vanacker G, & McEvoy L, et al.: Maintenance of graft compression in the adult cervical spine. Eur Spine J 15:1204–1209, 2006
Bridwell KH: Decision making regarding Smith-Petersen vs. pedicle subtraction osteotomy vs vertebral column resection for spinal deformity. Spine (Phila Pa 1976) 31:19 Suppl S171–S178, 2006
Briem D, , Windolf J, , Lehmann W, , Begemann PGC, & Meenen NM: [Bone grafts endoscopically applied to the spine. Results and therapeutic consequences.]. Unfallchirurg 107:1152–1161, 2004. (Ger)
Buchowski JM, , Bridwell KH, , Lenke LG, , Kuhns CA, , Lehman RA Jr, & Kim YJ, et al.: Neurologic complications of lumbar pedicle subtraction osteotomy: a 10-year assessment. Spine (Phila Pa 1976) 32:2245–2252, 2007
Buchowski JM, , Kuhns CA, , Bridwell KH, & Lenke LG: Surgical management of posttraumatic thoracolumbar kyphosis. Spine J 8:666–677, 2008
Carl AL, , Tranmer BI, & Sachs BL: Anterolateral dynamized instrumentation and fusion for unstable thoracolumbar and lumbar burst fractures. Spine (Phila Pa 1976) 22:686–690, 1997
Dai LY, , Chen WH, & Jiang LS: Anterior instrumentation for the treatment of pyogenic vertebral osteomyelitis of thoracic and lumbar spine. Eur Spine J 17:1027–1034, 2008
Dai LY, , Jiang LS, & Jiang SD: Anterior-only stabilization using plating with bone structural autograft versus titanium mesh cages for two- or three-column thoracolumbar burst fractures: a prospective randomized study. Spine (Phila Pa 1976) 34:1429–1435, 2009
Disch AC, , Knop C, , Schaser KD, , Blauth M, & Schmoelz W: Angular stable anterior plating following thoracolumbar corpectomy reveals superior segmental stability compared to conventional polyaxial plate fixation. Spine (Phila Pa 1976) 33:1429–1437, 2008
Erli HJ: First results with the anterior thoracolumbar reduction plate ArcoFix. Eur Spine J 18:1759, 2009. (Abstract)
Esses SI, , Botsford DJ, & Kostuik JP: Evaluation of surgical treatment for burst fractures. Spine (Phila Pa 1976) 15:667–673, 1990
Ferrera LA, , Gordon I, , Coquillette M, , Milks R, , Fleischman AJ, & Roy S, et al.: A preliminary biomechanical evaluation in a simulated spinal fusion model. Laboratory investigation. J Neurosurg Spine 7:542–548, 2007
Flamme CH, , von der Heide N, , Heymann C, & Hurschler C: Primary stability of anterior lumbar stabilization: interdependence of implant type and endplate retention or removal. Eur Spine J 15:807–818, 2006
Frank E, , Chamberland D, & Ragel B: A proposed technique for intraoperative measurement of cervical spine stiffness. Neurosurgery 39:147–150, 1996
Glassman SD, , Berven S, , Bridwell KH, , Horton W, & Dimar JR: Correlation of radiographic parameters and clinical symptoms in adult scoliosis. Spine (Phila Pa 1976) 30:682–688, 2005
Glassman SD, , Bridwell K, , Dimar JR, , Horton W, , Berven S, & Schwab F: The impact of positive sagittal balance in adult spinal deformity. Spine (Phila Pa 1976) 30:2024–2029, 2005
Goel VK, , Ramirez SA, , Kong W, & Gilbertson LG: Cancellous bone Young's modulus variation within the vertebral body of a ligamentous lumbar spine—application of bone adaptive remodeling concepts. J Biomech Eng 117:266–271, 1995
Goodship AE, , Cunningham JL, & Kenwright J: Strain rate and timing of stimulation in mechanical modulation of fracture healing. Clin Orthop Relat Res 355 Suppl S105–S115, 1998
Grant JP, , Oxland TR, & Dvorak MF: Mapping the structural properties of the lumbosacral vertebral endplates. Spine (Phila Pa 1976) 26:889–896, 2001
Ha KY, , Lee JS, & Kim KY: Bone graft volumetric changes and clinical outcomes after instrumented lumbar or lumbosacral fusion—a prospective cohort study with a 5-year follow-up. Eur Spine J 18:Suppl 4 S479–S480, 2009. (Abstract)
Hasegawa K, , Abe M, , Washio T, & Hara T: An experimental study on the interface strength between titanium mesh cage vertebra in reference to vertebral bone mineral density. Spine (Phila Pa 1976) 26:957–963, 2001
Hauck S, , Berger A, , Gonschorek O, & Bühren V: First results with a trabecular metal cage for monosegmental fusions in thoracoscopic fusion technique in unstable fractures of the thoracolumbar spine. Eur Spine J 17:1540–1633, 2008
Hauck S, , Gonschorek O, & Bühren V: First experience with a trabecular metal-cage in thoracoscopic monosegmental vertebral fracture treatment. Eur Spine J 18:1713–1826, 2009
Heary RF, & Bono CM: Pedicle subtraction osteotomy in the treatment of chronic, posttraumatic kyphotic deformity. J Neurosurg Spine 5:1–8, 2006
Hing KA: Bone repair in the twenty-first century: biology, chemistry or engineering?. Philos Transact A Math Phys Eng Sci 362:2821–2850, 2004
Hollowell JP, , Vollmer DG, , Wilson CR, , Pintar FA, & Yoganandan N: Biomechanical analysis of thoracolumbar interbody constructs. How important is the endplate?. Spine (Phila Pa 1976) 21:1032–1036, 1996
Huang P, , Gupta MC, , Sarigul-Klijn N, & Hazelwood S: Two in vivo surgical approaches for lumbar corpectomy using allograft and a metallic implant: a controlled clinical and biomechanical study. Spine J 6:648–658, 2006
Hussain M, , Natarajan RN, , Fayyazi AH, , Braaksma BR, , Andersson GBJ, & An HS: Screw angulation affects bone-screw stresses and bone graft load sharing in anterior cervical corpectomy fusion with a rigid screw-plate construct: a finite element model study. Spine J 9:1016–1023, 2009
Hwang SH, , Kayanja M, , Milks RA, & Benzel EC: Biomechanical comparison of adjacent segmental motion after ventral cervical fixation with varying angles of lordosis. Spine J 7:216–221, 2007
Illés T, , de Jonge T, , Domán I, & Dóczi T: Surgical correction of the late consequences of posttraumatic spinal disorders. J Spinal Disord Tech 15:127–132, 2002
Jin D, , Qu D, , Chen J, & Zhang H: One-stage anterior interbody autografting and instrumentation in primary surgical management of thoracolumbar spinal tuberculosis. Eur Spine J 13:114–121, 2004
Kaneda K, , Taneichi H, , Abumi K, , Hashimoto T, , Satoh S, & Fujiya M: Anterior decompression and stabilization with the Kaneda device for thoracolumbar burst fractures associated with neurological deficits. J Bone Joint Surg Am 79:69–83, 1997
Kawakami M, , Tamaki T, , Ando M, , Yamada H, , Hashizume H, & Yoshida M: Lumbar sagittal balance influences the clinical outcome after decompression and posterolateral spinal fusion for degenerative lumbar spondylolisthesis. Spine (Phila Pa 1976) 27:59–64, 2002
Kim SM, , Lim TJ, , Paterno J, , Park J, & Kim DH: Biomechanical comparison: stability of lateral-approach anterior lumbar interbody fusion and lateral fixation compared with anterior-approach anterior lumbar interbody fusion and posterior fixation in the lower lumbar spine. J Neurosurg Spine 2:62–68, 2005
Kin A, , Baba I, , Tsuboi K, , Kitahara Y, , Shiraishi M, & Murakoshi F, et al.: Vertebral endplate shape and measurement at fusion levels in degenerative lumbar disease. Eur Spine J 18:Suppl 4 S483, 2009. (Abstract)
Knop C, , Blauth M, , Bühren V, , Arand M, , Egbers HJ, & Hax PM, et al.: [Surgical treatment of injuries of the thoracolumbar transition--3: Follow-up examination. Results of a prospective multi-center study by the ”Spinal” Study Group of the German Society of Trauma Surgery.]. Unfallchirurg 104:583–600, 2001. (Ger)
Knop C, , Kranabetter T, , Reinhold M, & Blauth M: Combined posterior-anterior stabilisation of thoracolumbar injuries utilising a vertebral body replacing implant. Eur Spine J 18:949–963, 2009
Knop C, , Lange U, , Bastian L, , Oeser M, & Blauth M: Biomechanical compression tests with a new implant for thoracolumbar vertebral body replacement. Eur Spine J 10:30–37, 2001
Korovessis P, , Baikousis A, , Zacharatos S, , Petsinis G, , Koureas G, & Iliopoulos P: Combined anterior plus posterior stabilization versus posterior short-segment instrumentation and fusion for mid-lumbar (L2–L4) burst fractures. Spine (Phila Pa 1976) 31:859–868, 2006
Lafage V, , Schwab F, , Patel A, , Hawkinson N, & Farcy JP: Pelvic tilt and truncal inclination: two key radiographic parameters in the setting of adults with spinal deformity. Spine (Phila Pa 1976) 34:E599–E606, 2009
Lange U, , Edeling S, , Knop C, , Bastian L, , Oeser M, & Krettek C, et al.: Anterior vertebral body replacement with a titanium implant of adjustable height: a prospective clinical study. Eur Spine J 16:161–172, 2007
Le Huec JC, , Tournier C, , Aunoble S, , Madi K, & Leijssen P: Video-assisted treatment of thoracolumbar junction fractures using a specific distractor for reduction: prospective study of 50 cases. Eur Spine J 19:Suppl 1 S27–S32, 2009
Lerner T, , Schulter TL, , Vieth V, , Bullmann V, & Liljenqvist U: Subsidence rate and osseous integration of titanium ring cages for anterior spinal column reconstruction in vertebral osteomyelitis and malignant tumors. Eur Spine J 18:Suppl 4 S484, 2009. (Abstract)
Liljenqvist UR, , Bullmann V, , Schulte TL, , Hackenberg L, & Halm HF: Anterior dual rod instrumentation in idiopathic thoracic scoliosis. Eur Spine J 15:1118–1127, 2006
Lim TH, , An HS, , Hong JH, , Ahn JY, , You JW, & Eck J, et al.: Biomechanical evaluation of anterior and posterior fixations in an unstable calf spine model. Spine (Phila Pa 1976) 22:261–266, 1997
Matschke S, , Wagner C, , Davids D, & Wentzensen A: Complications in endoscopic anterior thoracolumbar spinal reconstructive surgery. Eur J Trauma 3:215–226, 2006
McAfee PC, , Farey ID, , Sutterlin CE, , Gurr KR, , Warden KE, & Cunningham BW: 1989 Volvo Award in basic science. Device-related osteoporosis with spinal instrumentation. Spine (Phila Pa 1976) 14:919–926, 1989
McDonough PW, , Davis R, , Tribus C, & Zdeblick TA: The management of acute thoracolumbar burst fractures with anterior corpectomy and Z-plate fixation. Spine (Phila Pa 1976) 29:1901–1909, 2004
Morrison RH, , Thierolf A, & Weckbach A: Volumetric changes of iliac crest autografts used to reconstruct the anterior column in thoracolumbar fractures: a follow-up using CT scans. Spine (Phila Pa 1976) 32:3030–3035, 2007
Nydegger T, , Oxland TR, , Hoffer Z, , Cottle W, & Nolte LP: Does anterolateral cage insertion enhance immediate stabilization of the functional spinal unit? A biomechanical investigation. Spine (Phila Pa 1976) 26:2491–2497, 2001
Oda I, , Cunningham BW, , Abumi K, , Kaneda K, & McAfee PC: The stability of reconstruction methods after thoracolumbar total spondylectomy. An in vitro investigation. Spine (Phila Pa 1976) 24:1634–1638, 1999
Oda T, & Panjabi MM: Pedicle screw adjustments affect stability of thoracolumbar burst fracture. Spine (Phila Pa 1976) 26:2328–2333, 2001
Okuyama K, , Abe E, , Chiba M, , Ishikawa N, & Sato K: Outcome of anterior decompression and stabilization for thoracolumbar unstable burst fractures in the absence of neurologic deficits. Spine (Phila Pa 1976) 21:620–625, 1996
Olsewski JM, , Garvey TA, & Schendel MJ: Biomechanical analysis of facet and graft loading in a Smith-Robinson type cervical spine model. Spine (Phila Pa 1976) 19:2540–2544, 1994
Oskouian RJ Jr, , Shaffrey CI, , Whitehill R, , Sansur CA, , Pouratian N, & Kanter AS, et al.: Anterior stabilization of three-column thoracolumbar spinal trauma. J Neurosurg Spine 5:18–25, 2006
Paley D: Principles of Deformity Correction Berlin, Springer-Verlag, 2005
Pearcy MJ, & Bogduk N: Instantaneous axes of rotation of the lumbar intervertebral joints. Spine (Phila Pa 1976) 13:1033–1041, 1988
Pilliar RM, , Lee JM, & Maniatopoulos C: Observations on the effect of movement on bone ingrowth into porous-surfaced implants. Clin Orthop Relat Res 208:108–113, 1986
Rauzzino MJ, , Shaffrey CI, , Nockels RP, , Wiggins GC, , Rock J, & Wagner J: Anterior lumbar fusion with titanium threaded and mesh interbody cages. Neurosurg Focus 7:6 e7, 1999
Reddi V, , Clarke DV Jr, & Arlet V: Anterior thoracoscopic instrumentation in adolescent idiopathic scoliosis: a systematic review. Spine (Phila Pa 1976) 33:1986–1994, 2008
Reinhold M, , Knop C, , Beisse R, , Audigé L, , Kandziora F, & Pizanis A, et al.: [Operative treatment of traumatic fractures of the thoracic and lumbar spinal column. Part I: epidemiology.]. Unfallchirurg 112:33–42, 2009. (Ger)
Reinhold M, , Schmölz W, , Canto F, , Krappinger D, , Blauth M, & Knop C: [An improved vertebral body replacement for the thoracolumbar spine. A biomechanical in vitro test on human lumbar vertebral bodies.]. Unfallchirurg 110:327–333, 2007. (Ger)
Robertson PA, , Rawlinson HJ, & Hadlow AT: Radiologic stability of titanium mesh cages for anterior spinal reconstruction following thoracolumbar corpectomy. J Spinal Disord Tech 17:44–52, 2004
Rohlmann A, , Bergmann G, , Graichen F, & Neff G: Braces do not reduce loads on internal spinal fixation devices. Clin Biomech (Bristol, Avon) 14:97–102, 1999
Rohlmann A, , Graichen F, , Bender A, , Kayser R, & Bergmann G: Loads on a telemeterized vertebral body replacement measured in three patients within the first postoperative month. Clin Biomech (Bristol, Avon) 23:147–158, 2008
Rohlmann A, , Graichen F, , Kayser R, , Bender A, & Bergmann G: Loads on a telemeterized vertebral body replacement measured in two patients. Spine (Phila Pa 1976) 33:1170–1179, 2008
Rohlmann A, , Zander T, , Fehrmann M, , Klöckner C, & Bergmann G: [Effect of implants for replacement of a vertebral body on the mechanical function of the lumbar spine.]. Orthopade 31:503–507, 2002. (Ger)
Sasso RC, , Renkens K, , Hanson D, , Reilly T, , McGuire RA Jr, & Best NM: Unstable thoracolumbar burst fractures: anterioronly versus short-segment posterior fixation. J Spinal Disord Tech 19:242–248, 2006
Schultheiss M, , Kinzl L, , Claes L, , Wilke HJ, & Hartwig E: Minimally invasive ventral spondylodesis for thoracolumbar fracture treatment: surgical technique and first clinical outcome. Eur Spine J 12:618–624, 2003
Stoltze D, & Harms J: [Correction of posttraumatic deformities. Principles and methods.]. Orthopade 28:731–745, 1999. (Ger)
Tayyab NA, , Samartzis D, , Altiok H, , Shuff CE, , Lubicky JP, & Hermann J, et al.: The reliability and diagnostic value of radiographic criteria in sagittal spine deformities: comparison of the vertebral wedge ratio to the segmental cobb angle. Spine (Phila Pa 1976) 32:E451–E459, 2007
Truumees E, , Demetropoulos CK, , Yang KH, & Herkowitz HN: Effects of a cervical compression plate on graft forces in an anterior cervical discectomy model. Spine (Phila Pa 1976) 28:1097–1102, 2003
Truumees E, , Demetropoulos CK, , Yang KH, & Herkowitz HN: Effects of disc height and distractive forces on graft compression in an anterior cervical corpectomy model. Spine (Phila Pa 1976) 33:1438–1441, 2008
Truumees E, , Demetropoulos CK, , Yang KH, & Herkowitz HN: Effects of disc height and distractive forces on graft compression in an anterior cervical discectomy model. Spine (Phila Pa 1976) 27:2441–2445, 2002
Wang JL, , Panjabi MM, & Isomi T: The role of bone graft force in stabilizing the multilevel anterior cervical spine plate system. Spine (Phila Pa 1976) 25:1649–1654, 2000
Wiggins GC, , Rauzzino MJ, , Shaffrey CI, , Nockels RP, , Whitehill R, & Alden TD, et al.: A new technique for the surgical management of unstable thoracolumbar burst fractures: a modification of the anterior approach and an outcome comparison to traditional methods. Neurosurg Focus 7:1 e3, 1999
Wilke HJ, , Jungkunz B, , Wenger K, & Claes LE: Spinal segment range of motion as a function of in vitro test conditions: effects of exposure period, accumulated cycles, angular-deformation rate, and moisture condition. Anat Rec 251:15–19, 1998
Wilke HJ, , Wenger K, & Claes L: Testing criteria for spinal implants: recommendations for the standardization of in vitro stability testing of spinal implants. Eur Spine J 7:148–154, 1998
Wilson JA, , Bowen S, , Branch CL Jr, & Meredith JW: Review of 31 cases of anterior thoracolumbar fixation with the anterior thoracolumbar locking plate system. Neurosurg Focus 7:1 e1, 1999
Wood KB, , Bohn D, & Mehbod A: Anterior versus posterior treatment of stable thoracolumbar burst fractures without neurologic deficit: a prospective, randomized study. J Spinal Disord Tech 18:Suppl S15–S23, 2005
Yoshioka T, , Tsuji H, , Hirano N, & Sainoh S: Motion characteristic of the normal lumbar spine in young adults: instantaneous axis of rotation and vertebral center motion analyses. J Spinal Disord 3:103–113, 1990
All Time | Past Year | Past 30 Days | |
---|---|---|---|
Abstract Views | 540 | 64 | 3 |
Full Text Views | 162 | 12 | 0 |
PDF Downloads | 315 | 8 | 0 |
EPUB Downloads | 0 | 0 | 0 |