Object

The authors describe a method of using computer models to generate customized cervical implants. A promising yet challenging technique in cervical spine surgery involves the use of pedicle screws to assist with posterior instrumentation. Surrounding anatomical structures such as the vertebral arteries and cervical nerve roots present challenges for safe screw placement; however, the use of computer-generated templates seems to be a promising method to assist with placement. In this study, the authors explore the use of computer-generated templates and introduce their methods for creating custom, bioabsorbable posterior cervical implants.

Methods

The cervical spines (C2–T1) from 4 cadavers were scanned with volumetric CT. Using commercially available software, the authors generated volumetric models of a cervical drill template and the mold for a cervical plate spanning a desired number of vertebrae. The computer generated models of the cervical drill template and cervical plate mold were converted into physical models using a rapid prototyping machine. The biopolymer polylactic acid resin was heated to 250°C and resolidified to form thin ~ 5-mm-thick plates. The newly formed plates were reheated to 60°C and cast on the cervical mold.

Results

The resulting translucent plates were found on visual inspection to have a secure lock-and-key fit on the original cadaver spine, and the techniques used were robust and reproducible. The process described in this brief report provides the background to proceed with development and testing of these patient-absorbable templates.

Conclusions

The creation and use of patient-specific bioabsorbable posterior cervical plates in conjunction with multilevel drill templates appear promising. Additional feasibility studies are planned, and in vitro studies are required to determine the safety and efficacy of using patient-specific drill templates and converting them into bioabsorbable implants.

Abbreviations used in this paper: ABS = acrylonitrile butadiene styrene; PLA = polylactic acid; PS = pedicle screw; STL = stereolithography.

Spine - 1 year subscription bundle (Individuals Only)

USD  $369.00

JNS + Pediatrics + Spine - 1 year subscription bundle (Individuals Only)

USD  $600.00

Contributor Notes

Address correspondence to: Timothy C. Ryken, M.D., Department of Neurosurgery, University of Iowa Carver College of Medicine, 1846 JPP 200 Hawkins Drive, Iowa City, Iowa 52242. email: timothy-ryken@uiowa.edu.
  • 1

    Ames CP, , Acosta FL Jr, , Chamberlain RH, , Larios AE, & Crawford NR: Biomechanical analysis of a newly designed bioabsorbable anterior cervical plate. J Neurosurg Spine 3:465470, 2005

    • Search Google Scholar
    • Export Citation
  • 2

    Berry E, , Cuppone M, , Porada S, , Millner PA, , Rao A, & Chiverton N, : Personalised image-based templates for intra-operative guidance. Proc Inst Mech Eng [H] 219:111118, 2005

    • Search Google Scholar
    • Export Citation
  • 3

    Claes LE, , Ignatius AA, , Rehm KE, & Scholz C: New bioresorbable pin for the reduction of small bony fragments: design, mechanical properties and in vitro degradation. Biomaterials 17:16211626, 1996

    • Search Google Scholar
    • Export Citation
  • 4

    Goffin J, , Van Brussel K, , Martens K, , Vander Sloten J, , Van Audekercke R, & Smet MH: Three-dimensional computed tomography-based, personalized drill guide for posterior cervical stabilization at C1-C2. Spine 26:13431347, 2001

    • Search Google Scholar
    • Export Citation
  • 5

    Gunatillake PA, & Adhikari R: Biodegradable synthetic polymers for tissue engineering. Eur Cell Mater 5:116, 2003

  • 6

    Hollinger JO: Preliminary report on osteogenic potential of a biodegradable copolymer of polylactide (PLA) and polyglycolide (PGA). J Biomed Mater Res 17:7182, 1983

    • Search Google Scholar
    • Export Citation
  • 7

    Jones EL, , Heller JG, , Silcox DH, & Hutton WC: Cervical pedicle screws versus lateral mass screws. Anatomic feasibility and biomechanical comparison. Spine 22:977982, 1997

    • Search Google Scholar
    • Export Citation
  • 8

    Kast E, , Mohr K, , Richter HP, & Borm W: Complications of transpedicular screw fixation in the cervical spine. Eur Spine J 15:327334, 2005

    • Search Google Scholar
    • Export Citation
  • 9

    Ludwig SC, , Kramer DL, , Vaccaro AR, & Albert TJ: Transpedicle screw fixation of the cervical spine. Clin Orthop Relat Res 359:7788, 1999

  • 10

    Mac-Thiong JM, , Labelle H, , Rooze M, , Feipel V, & Aubin CE: Evaluation of a transpedicular drill guide for pedicle screw placement in the thoracic spine. Eur Spine J 12:542547, 2003

    • Search Google Scholar
    • Export Citation
  • 11

    Nelson JF, , Stanford HG, & Cutright DE: Evaluation and comparison of biodegradable substances as osteogenic agents. Oral Surg Oral Med Oral Pathol 43:836843, 1977

    • Search Google Scholar
    • Export Citation
  • 12

    Neo M, , Fujibayashi S, , Yoshida M, & Nakamura T: Spinous process plate fixation as a salvage operation for failed anterior cervical fusion. Technical note. J Neurosurg Spine 4:7881, 2006

    • Search Google Scholar
    • Export Citation
  • 13

    Owen BD, , Christensen GE, , Reinhardt JM, & Ryken TC: Rapid prototype patient-specific drill template for cervical pedicle screw ploacement. Comput Aided Surg 12:303308, 2007

    • Search Google Scholar
    • Export Citation
  • 14

    Radermacher K, , Portheine F, , Anton M, , Zimolong A, , Kaspers G, & Rau G, : Computer assisted orthopaedic surgery with image based individual templates. Clin Orthop Relat Res 354:2838, 1998

    • Search Google Scholar
    • Export Citation
  • 15

    Rezcallah AT, , Xu R, , Ebraheim NA, & Jackson T: Axial computed tomography of the pedicle in the lower cervical spine. Am J Orthop 30:5961, 2001

    • Search Google Scholar
    • Export Citation
  • 16

    Steinmann JC, , Herkowitz HN, , el-Kommos H, & Wesolowski DP: Spinal pedicle fixation. Confirmation of an image-based technique for screw placement. Spine 18:18561861, 1993

    • Search Google Scholar
    • Export Citation
  • 17

    Yoshimoto H, , Sato S, , Hyakumachi T, , Yanagibashi Y, & Masuda T: Spinal reconstruction using a cervical pedicle screw system. Clin Orthop Relat Res 431:111119, 2005

    • Search Google Scholar
    • Export Citation

Metrics

All Time Past Year Past 30 Days
Abstract Views 512 95 6
Full Text Views 68 4 0
PDF Downloads 121 4 0
EPUB Downloads 0 0 0