Mark Nowell, Rachel Sparks, Gergely Zombori, Anna Miserocchi, Roman Rodionov, Beate Diehl, Tim Wehner, Gianluca Baio, Gianluca Trevisi, Martin Tisdall, Sebastien Ourselin, Andrew W. McEvoy and John Duncan

OBJECT

The objective of this study was to evaluate the clinical utility of multitrajectory computer-assisted planning software (CAP) to plan stereoelectroencephalography (SEEG) electrode arrangements.

METHODS

A cohort of 18 patients underwent SEEG for evaluation of epilepsy at a single center between August 2013 and August 2014. Planning of electrodes was performed manually and stored using EpiNav software. CAP was developed as a planning tool in EpiNav. The user preselects a set of cerebral targets and optimized trajectory constraints, and then runs an automated search of potential scalp entry points and associated trajectories. Each trajectory is associated with metrics for a safety profile, derived from the minimal distance to vascular structures, and an efficacy profile, derived from the proportion of depth electrodes that are within or adjacent to gray matter. CAP was applied to the cerebral targets used in the cohort of 18 previous manually planned implantations to generate new multitrajectory implantation plans. A comparison was then undertaken for trajectory safety and efficacy.

RESULTS

CAP was applied to 166 electrode targets in 18 patients. There were significant improvements in both the safety profile and efficacy profile of trajectories generated by CAP compared with manual planning (p < 0.05). Three independent neurosurgeons assessed the feasibility of the trajectories generated by CAP, with 131 (78.9%) of 166 trajectories deemed suitable for implementation in clinical practice. CAP was performed in real time, with a median duration of 8 minutes for each patient, although this does not include the time taken for data preparation.

CONCLUSIONS

CAP is a promising tool to plan SEEG implantations. CAP provides feasible depth electrode arrangements, with quantitatively greater safety and efficacy profiles, and with a substantial reduction in duration of planning within the 3D multimodality framework.

Restricted access

Roman Rodionov, Aidan O’Keeffe, Mark Nowell, Michele Rizzi, Vejay N. Vakharia, Victoria Wykes, Sofia H. Eriksson, Anna Miserocchi, Andrew W. McEvoy, Sebastien Ourselin and John S. Duncan

OBJECTIVE

The accuracy of stereoelectroencephalography (SEEG) electrode implantation is an important factor in maximizing its safety. The authors established a quality assurance (QA) process to aid advances in implantation accuracy.

METHODS

The accuracy of three consecutive modifications of a frameless implantation technique was quantified in three cohorts comprising 22, 8, and 23 consecutive patients. The modifications of the technique aimed to increase accuracy of the bolt placement.

RESULTS

The lateral shift of the axis of the implanted bolt at the level of the planned entry point was reduced from a mean of 3.0 ± 1.6 mm to 1.4 ± 0.8 mm. The lateral shift of the axis of the implanted bolt at the level of the planned target point was reduced from a mean of 3.8 ± 2.5 mm to 1.6 ± 0.9 mm.

CONCLUSIONS

This QA framework helped to isolate and quantify the factors introducing inaccuracy in SEEG implantation, and to monitor ongoing accuracy and the effect of technique modifications.