Laura A. Snyder, Jennifer N. Lehrman, Ram Kumar Menon, Jakub Godzik, Anna G. U. S. Newcomb and Brian P. Kelly

OBJECTIVE

Minimally invasive transforaminal interbody fusion techniques vary among surgeons. One decision point is whether to perform a unilateral facetectomy (UF), a unilateral facetectomy plus partial contralateral facetectomy (UF/PF), or a complete bilateral facetectomy (CBF). The authors therefore compared the biomechanical benefits of all 3 types of facetectomies to determine which approach produces improved biomechanical outcomes.

METHODS

Seven human cadaveric specimens (L3–S1) were potted and prepped for UF, with full facet removal, hemilaminectomy, discectomy, and pedicle screw placement. After distraction, a fixed interbody spacer was placed, and compression was performed. A final fixation configuration was performed by locking the rods across the screws posteriorly with bilateral compression. Final lordosis angle and change and foraminal height were measured, and standard nondestructive flexibility tests were performed to assess intervertebral range of motion (ROM) and compressive stiffness. The same procedure was followed for UF/PF and CBF in all 7 specimens.

RESULTS

All 3 conditions demonstrated similar ROM and compressive stiffness. No statistically significant differences occurred with distraction, but CBF demonstrated significantly greater change than UF in mean foraminal height after bilateral posterior compression (1.90 ± 0.62 vs 1.00 ± 0.45 mm, respectively, p = 0.04). With compression, the CBF demonstrated significantly greater mean ROM than the UF (2.82° ± 0.83° vs 2.170° ± 1.10°, p = 0.007). The final lordosis angle was greatest with CBF (3.74° ± 0.70°) and lowest with UF (2.68° ± 1.28°). This finding was statistically significant across all 3 conditions (p ≤ 0.04).

CONCLUSIONS

Although UF/PF and CBF may require slightly more time and effort and incur more risk than UF, the potential improvement in sagittal balance may be worthwhile for select patients.

Restricted access

Jakub Godzik, Jennifer N. Lehrman, Anna G. U. S. Newcomb, Ram Kumar Menon, Alexander C. Whiting, Brian P. Kelly and Laura A. Snyder

OBJECTIVE

Transforaminal lumbar interbody fusion (TLIF) is commonly used for lumbar fusion, such as for foraminal decompression, stabilization, and improving segmental lordosis. Although many options exist, surgical success is contingent on matching design strengths with surgical goals. The goal in the present study was to investigate the effects of an expandable interbody spacer and 2 traditional static spacer designs in terms of stability, compressive stiffness, foraminal height, and segmental lordosis.

METHODS

Standard nondestructive flexibility tests (7.5 N⋅m) were performed on 8 cadaveric lumbar specimens (L3–S1) to assess intervertebral stability of 3 types of TLIF spacers at L4–5 with bilateral posterior screw-rod (PSR) fixation. Stability was determined as range of motion (ROM) in flexion-extension (FE), lateral bending (LB), and axial rotation (AR). Compressive stiffness was determined with axial compressive loading (300 N). Foraminal height, disc height, and segmental lordosis were evaluated using radiographic analysis after controlled PSR compression (170 N). Four conditions were tested in random order: 1) intact, 2) expandable interbody cage with PSR fixation (EC+PSR), 3) static ovoid cage with PSR fixation (SOC+PSR), and 4) static rectangular cage with PSR fixation (SRC+PSR).

RESULTS

All constructs demonstrated greater stability than the intact condition (p < 0.001). No significant differences existed among constructs in ROM (FE, AR, and LB) or compressive stiffness (p ≥ 0.66). The EC+PSR demonstrated significantly greater foraminal height at L4–5 than SRC+PSR (21.1 ± 2.6 mm vs 18.6 ± 1.7 mm, p = 0.009). EC+PSR demonstrated higher anterior disc height than SOC+PSR (14.9 ± 1.9 mm vs 13.6 ± 2.2 mm, p = 0.04) and higher posterior disc height than the intact condition (9.4 ± 1.5 mm vs 7.1 ± 1.0 mm, p = 0.002), SOC+PSR (6.5 ± 1.8 mm, p < 0.001), and SRC+PSR (7.2 ± 1.2 mm, p < 0.001). There were no significant differences in segmental lordosis among SOC+PSR (10.1° ± 2.2°), EC+PSR (8.1° ± 0.5°), and SRC+PSR (11.1° ± 3.0°) (p ≥ 0.06).

CONCLUSIONS

An expandable interbody spacer provided stability, stiffness, and segmental lordosis comparable to those of traditional nonexpandable spacers of different shapes, with increased foraminal height and greater disc height. These results may help inform decisions about which interbody implants will best achieve surgical goals.