Disparity between adjusted and actual opening cerebrospinal fluid pressure in a patient with the Codman Hakim programmable valve: occult form of shunt failure due to head banging

Case report

Restricted access

✓ Although the Codman Hakim programmable valve (CHPV) system is popular, several problems arising from its design have been described. The authors report an additional cause of shunt failure associated with the use of the CHPV system. A 7-year-old girl who had previously undergone placement of a ventriculoperitoneal shunt with the CHPV system was admitted in a comatose state. Radiography did not demonstrate any breakage or change in the opening pressure of the valve. Manual pumping of the valve resulted in neurological improvement and shrinkage of the girl’s ventricles. However, ventriculomegaly was again observed several hours later, and an emergency valve revision was performed.

The manufacturer’s examination of the extracted valve revealed a crack on the surface of the hard plastic housing covering the valve chamber, which appeared to be suggestive of some type of blunt trauma. In addition, the flat spring that transmitted resistance from the pressure control cam to the valve ball was deformed in such a manner that it caused excessive pressure against the valve ball. Because of this deformity, the actual opening pressure in this patient had reached a maximum level of 226 mm H2O instead of the adjusted pressure of 60 mm H2O. In patients with the CHPV system whose medical history is difficult to ascertain, the possibility of damage to the valve spring and the need for a prompt revision surgery should be considered. The authors conclude that placement of the CHPV system may be contraindicated in patients with a known habit of head banging.

Abbreviations used in this paper:CHPV = Codman Hakim programmable valve; CSF = cerebrospinal fluid; CT = computed tomography.
Article Information

Contributor Notes

Address reprint requests to: Kimitoshi Sato, M.D., Department of Neurosurgery, Kitasato University School of Medicine, 1-15-1 Ki-tasato, Sagamihara, Kanagawa 228-8555, Japan. email: kimitoshi@k8.dion.ne.jp.

© AANS, except where prohibited by US copyright law.

Headings
References
  • 1

    Belliard HRoux FXTurak BNataf FDevaux BCioloca C: [The Codman Medos programmable shunt valve. Evaluation of 53 implantations in 50 patients.]. Neurochirurgie 42:1391451996. (Fr)

    • Search Google Scholar
    • Export Citation
  • 2

    Black PMHakim RBailey NO: The use of the Codman-Medos Programmable Hakim valve in the management of patients with hydrocephalus: illustrative cases. Neurosurgery 34:111011131994

    • Search Google Scholar
    • Export Citation
  • 3

    Drake JMKestle JRWMilner RCinalli GBoop FPiatt J Jr: Randomized trial of cerebrospinal fluid shunt valve design in pediatric hydrocephalus. Neurosurgery 43:2943031998

    • Search Google Scholar
    • Export Citation
  • 4

    Hakim C: The Physics and Pathophysiology of the Hydraulic Complex of the Central Nervous System. The Mechanics of Hydrocephalus and Normal Pressure Hydrocephalus. PhD Thesis BostonMassachusetts Institute of Technology1985

    • Search Google Scholar
    • Export Citation
  • 5

    Hakim SDuran de la Roche FBurton JD: A critical analysis of valve shunts used in the treatment of hydrocephalus. Dev Med Child Neurol 15:2302551973

    • Search Google Scholar
    • Export Citation
  • 6

    Kurosaki KHamada HHayashi NKurimoto MHirashima YEndo S: A rare case of shunt malfunction attributable to blockage of a Codman-Hakim programmable shunt valve. Childs Nerv Syst 18:1831852002

    • Search Google Scholar
    • Export Citation
  • 7

    Nabika SOki SSumida MIsobe N: Shunt malfunction caused by mischievous twisting of the valve by the patient: case report. Neurosurgery 53:2382402003

    • Search Google Scholar
    • Export Citation
  • 8

    Okazaki TOki SMigita KKurisu K: A rare case of shunt malfunction attributable to a broken Codman-Hakim programmable shunt valve after a blow to the head. Pediatr Neurosurg 41:2412432005

    • Search Google Scholar
    • Export Citation
  • 9

    Reinprecht ADietrich WBertalanffy ACzech T: The Medos Hakim programmable valve in the treatment of pediatric hydrocephalus. Childs Nerv Syst 13:5885931997

    • Search Google Scholar
    • Export Citation
  • 10

    Shimizu SUtsuki SSuzuki SOka HFujii K: Obstruction of a Codman-Hakim programmable valve by a migrating pressure control cam. J Neurosurg 103:3 Suppl2702712005

    • Search Google Scholar
    • Export Citation
  • 11

    Yamashita NKamiya KYamada K: Experience with a programmable valve shunt system. J Neurosurg 91:26311999

  • 12

    Zemack GRomner B: Seven years of clinical experience with the programmable Codman Hakim valve: a retrospective study of 583 patients. J Neurosurg 92:9419482000

    • Search Google Scholar
    • Export Citation
TrendMD
Metrics

Metrics

All Time Past Year Past 30 Days
Abstract Views 134 131 36
Full Text Views 47 38 0
PDF Downloads 35 23 0
EPUB Downloads 0 0 0
PubMed