The baric probe: a novel long-term implantable intracranial pressure monitor with ultrasound-based interrogation

Technical note

Restricted access

Object

Prompt diagnosis of shunt malfunction is critical in preventing neurological morbidity and death in individuals with hydrocephalus; however, diagnostic methods for this condition remain limited. For several decades, investigators have sought a long-term, implantable intracranial pressure (ICP) monitor to assist in the diagnosis of shunt malfunction, but efforts have been impeded by device complexity, marked measurement drift, and limited instrumentation lifespan. In the current report, the authors introduce an entirely novel, simple, compressible gas design that addresses each of these problems.

Methods

The device described herein, termed the “baric probe,” consists of a subdural fluid bladder and multichannel indicator that monitors the position of an air-fluid interface (AFI). A handheld ultrasound probe is used to interrogate the baric probe in vivo, permitting noninvasive ICP determination. To assess the function of device prototypes, ex vivo experiments were conducted using a water column, and short- and long-term in vivo experiments were performed using a porcine model with concurrent measurements of ICP via a fiberoptic monitor.

Results

Following a toe region of approximately 2 cm H2O, the baric probe's AFI demonstrated a predictable linear relationship to ICP in both ex vivo and in vivo models. After a 2-week implantation of the device, this linear relationship remained robust and reproducible. Further, changes in ICP were observed with the baric probe, on average, 3 seconds in advance of the fiberoptic ICP monitor reading.

Conclusions

The authors demonstrate “proof-of-concept” and feasibility for the baric probe, a long-term implantable ICP monitor designed to facilitate the prompt and accurate diagnosis of shunt malfunction. The baric probe showed a consistent linear relationship between ICP and the device's AFI in ex vivo and short- and long-term in vivo models. With a low per-unit cost, a reduced need for radiography or CT, and an indicator that can be read with a handheld ultrasound probe that interfaces with any smart phone, the baric probe promises to simplify the care of patients with shunt-treated hydrocephalus throughout both the developed and the developing world.

Abbreviations used in this paper:AFI = air-fluid interface; ICP = intracranial pressure.
Article Information

Contributor Notes

Address correspondence to: David D. Limbrick Jr., M.D., Ph.D., Department of Neurological Surgery, Washington University School of Medicine, St. Louis Children's Hospital, One Children's Place, Suite 4S20, St. Louis, Missouri 63110-1077. email: limbrickd@wudosis.wustl.edu.Please include this information when citing this paper: published online September 28, 2012; DOI: 10.3171/2012.8.PEDS11565.

© AANS, except where prohibited by US copyright law.

Headings
References
  • 1

    Aquilina KThoresen MChakkarapani EPople IKCoakham HBEdwards RJ: Preliminary evaluation of a novel intraparenchymal capacitive intracranial pressure monitor. Laboratory investigation. J Neurosurg 115:5615692011

    • Search Google Scholar
    • Export Citation
  • 2

    Atkinson JRShurtleff DBFoltz EL: Radio telemetry for the measurement of intracranial pressure. J Neurosurg 27:4284321967

  • 3

    Berrington de González AMahesh MKim KPBhargavan MLewis RMettler F: Projected cancer risks from computed tomographic scans performed in the United States in 2007. Arch Intern Med 169:207120772009

    • Search Google Scholar
    • Export Citation
  • 4

    Chern JJMacias CGJea ACurry DJLuerssen TGWhitehead WE: Effectiveness of a clinical pathway for patients with cerebrospinal fluid shunt malfunction. Clinical article. J Neurosurg Pediatr 6:3183242010

    • Search Google Scholar
    • Export Citation
  • 5

    Chou SYDigre KB: Neuro-ophthalmic complications of raised intracranial pressure, hydrocephalus, and shunt malfunction. Neurosurg Clin N Am 10:5876081999

    • Search Google Scholar
    • Export Citation
  • 6

    Committee for the Update of the Guide for the Care and Use of Laboratory Animals: Guide for the Care and Use of Laboratory Animals ed 8Washington, DCNational Academies Press2011

    • Search Google Scholar
    • Export Citation
  • 7

    Iskandar BJTubbs SMapstone TBGrabb PABartolucci AAOakes WJ: Death in shunted hydrocephalic children in the 1990s. Pediatr Neurosurg 28:1731761998

    • Search Google Scholar
    • Export Citation
  • 8

    Kulkarni AVDrake JMLamberti-Pasculli M: Cerebrospinal fluid shunt infection: a prospective study of risk factors. J Neurosurg 94:1952012001

    • Search Google Scholar
    • Export Citation
  • 9

    Miller JPFulop SCDashti SRRobinson SCohen AR: Rethinking the indications for the ventriculoperitoneal shunt tap. J Neurosurg Pediatr 1:4354382008

    • Search Google Scholar
    • Export Citation
  • 10

    Noetzel MJBaker RP: Shunt fluid examination: risks and benefits in the evaluation of shunt malfunction and infection. J Neurosurg 61:3283321984

    • Search Google Scholar
    • Export Citation
  • 11

    Patwardhan RVNanda A: Implanted ventricular shunts in the United States: the billion-dollar-a-year cost of hydrocephalus treatment. Neurosurgery 56:1391452005

    • Search Google Scholar
    • Export Citation
  • 12

    Paulsen AHLundar TLindegaard KF: Twenty-year outcome in young adults with childhood hydrocephalus: assessment of surgical outcome, work participation, and health-related quality of life. Clinical article. J Neurosurg Pediatr 6:5275352010

    • Search Google Scholar
    • Export Citation
  • 13

    Rocque BGLapsiwala SIskandar BJ: Ventricular shunt tap as a predictor of proximal shunt malfunction in children: a prospective study. Clinical article. J Neurosurg Pediatr 1:4394432008

    • Search Google Scholar
    • Export Citation
  • 14

    Sood SKim SHam SDCanady AIGreninger N: Useful components of the shunt tap test for evaluation of shunt malfunction. Childs Nerv Syst 9:1571621993

    • Search Google Scholar
    • Export Citation
  • 15

    Tuli SDrake JLawless JWigg MLamberti-Pasculli M: Risk factors for repeated cerebrospinal shunt failures in pediatric patients with hydrocephalus. J Neurosurg 92:31382000

    • Search Google Scholar
    • Export Citation
  • 16

    Welschehold SSchmalhausen EDodier PVulcu SOertel JWagner W: First clinical results with a new telemetric intracranial pressure-monitoring system. Neurosurgery 70:1 Suppl Operative44492012

    • Search Google Scholar
    • Export Citation
  • 17

    Winston KRLopez JAFreeman J: CSF shunt failure with stable normal ventricular size. Pediatr Neurosurg 42:1511552006

TrendMD
Metrics

Metrics

All Time Past Year Past 30 Days
Abstract Views 172 171 13
Full Text Views 86 72 3
PDF Downloads 58 35 2
EPUB Downloads 0 0 0
PubMed