Change in the angle of the anterior genu is associated with occlusion rate after Pipeline flow diversion

Al-Wala Awad MD, Craig Kilburg MD, Michael Karsy MD, PhD, MSc, William T. Couldwell MD, PhD, and Philipp Taussky MD
View More View Less
  • Department of Neurosurgery, Clinical Neurosciences Center, University of Utah, Salt Lake City, Utah
Restricted access

Purchase Now

USD  $45.00

JNS + Pediatrics - 1 year subscription bundle (Individuals Only)

USD  $505.00

JNS + Pediatrics + Spine - 1 year subscription bundle (Individuals Only)

USD  $600.00
Print or Print + Online

OBJECTIVE

The Pipeline embolization device (PED) is a self-expanding mesh stent that diverts blood flow away from an aneurysm; it has been successfully used to treat aneurysms of the proximal internal carotid artery (ICA). PEDs have a remarkable ability to alter regional blood flow along the tortuous segments of the ICA and were incidentally found to alter the angle of the anterior genu after treatment. The authors quantified these changes and explored their implications as they relate to treatment effect.

METHODS

The authors retrospectively reviewed cases of aneurysms treated with a PED between the ophthalmic and posterior communicating arteries from 2012 through 2015. The angles of the anterior genu were measured on the lateral projections of cerebral angiograms obtained before and after treatment with a PED. The angles of the anterior genu of patients without aneurysms were used as normal controls.

RESULTS

Thirty-eight patients were identified who had been treated with a PED; 34 (89.5%) had complete obliteration and 4 (10.5%) had persistence of their aneurysm at last follow-up (mean 11.3 months). After treatment, 32 patients had an increase, 3 had a decrease, and 3 had no change in the angle of the anterior genu. The average measured angle of the anterior genu was 36.7° before treatment and 44.3° after treatment (p < 0.0001). The average angle of the anterior genu of control patients was 43.32° (vs 36.7° for the preoperative angle in the patients with aneurysms, p < 0.057). The average change in the angle of patients with postoperative Raymond scores of 1 was 9.10°, as compared with 1.25° in patients with postoperative Raymond scores > 1 (p < 0.001).

CONCLUSIONS

Treatment with a PED significantly changes the angle of the anterior genu. An average change of 9.1° was associated with complete obliteration of treated aneurysms. These findings have important implications for the treatment and management of cerebral aneurysm.

ABBREVIATIONS

ICA = internal carotid artery; PED = Pipeline embolization device.

Figure from Minchev et al. (pp 150–158).

JNS + Pediatrics - 1 year subscription bundle (Individuals Only)

USD  $505.00

JNS + Pediatrics + Spine - 1 year subscription bundle (Individuals Only)

USD  $600.00

Contributor Notes

Correspondence Philipp Taussky: Clinical Neurosciences Center, University of Utah, Salt Lake City, UT. neuropub@hsc.utah.edu.

INCLUDE WHEN CITING Published online January 18, 2019; DOI: 10.3171/2018.9.JNS18726.

Disclosures Dr. Taussky reports a consultant relationship with Medtronic.

  • 1

    Becske T, Brinjikji W, Potts MB, Kallmes DF, Shapiro M, Moran CJ, et al. : Long-term clinical and angiographic outcomes following Pipeline Embolization Device treatment of complex internal carotid artery aneurysms: five-year results of the Pipeline for Uncoilable or Failed Aneurysms Trial. Neurosurgery 80:4048, 2017

    • Search Google Scholar
    • Export Citation
  • 2

    Becske T, Kallmes DF, Saatci I, McDougall CG, Szikora I, Lanzino G, et al. : Pipeline for uncoilable or failed aneurysms: results from a multicenter clinical trial. Radiology 267:858868, 2013

    • Search Google Scholar
    • Export Citation
  • 3

    Bouthillier A, van Loveren HR, Keller JT: Segments of the internal carotid artery: a new classification. Neurosurgery 38:425433, 1996

    • Search Google Scholar
    • Export Citation
  • 4

    Dai D, Ding YH, Kelly M, Kadirvel R, Kallmes D: Histopathological findings following pipeline embolization in a human cerebral aneurysm at the basilar tip. Interv Neuroradiol 22:153157, 2016

    • Search Google Scholar
    • Export Citation
  • 5

    Fischer S, Vajda Z, Aguilar Perez M, Schmid E, Hopf N, Bäzner H, et al. : Pipeline embolization device (PED) for neurovascular reconstruction: initial experience in the treatment of 101 intracranial aneurysms and dissections. Neuroradiology 54:369382, 2012

    • Search Google Scholar
    • Export Citation
  • 6

    Ishii A, Chihara H, Kikuchi T, Arai D, Ikeda H, Miyamoto S: Contribution of the straightening effect of the parent artery to decreased recanalization in stent-assisted coiling of large aneurysms. J Neurosurg 127:10631069, 2017

    • Search Google Scholar
    • Export Citation
  • 7

    Lin LM, Colby GP, Jiang B, Uwandu C, Huang J, Tamargo RJ, et al. : Classification of cavernous internal carotid artery tortuosity: a predictor of procedural complexity in Pipeline embolization. J Neurointerv Surg 7:628633, 2015

    • Search Google Scholar
    • Export Citation
  • 8

    Silva Neto ÂR, Câmara RL, Valença MM: Carotid siphon geometry and variants of the circle of Willis in the origin of carotid aneurysms. Arq Neuropsiquiatr 70:917921, 2012

    • Search Google Scholar
    • Export Citation
  • 9

    Suzuki T, Takao H, Fujimura S, Dahmani C, Ishibashi T, Mamori H, et al. : Relationships between geometrical parameters and mechanical properties for a helical braided flow diverter stent. Technol Health Care 25:611623, 2017

    • Search Google Scholar
    • Export Citation
  • 10

    Zou Q, Xue W, Lin J, Fu Y, Guan G, Wang F, et al. : Mechanical characteristics of novel polyester/NiTi wires braided composite stent for the medical application. Results Phys 6:440446, 2016

    • Search Google Scholar
    • Export Citation

Metrics

All Time Past Year Past 30 Days
Abstract Views 496 248 18
Full Text Views 140 70 3
PDF Downloads 141 39 3
EPUB Downloads 0 0 0